RF Leakage in Electrosurgical
RF leakage is radio frequency energy that flows from the electrosurgical generator
to earth ground rather than returning to the generator as part of the therapeutic path
(generator active accessory patient return electrode
generator). Theoretically, all therapeutic current delivered by the generator should
return to the generator via the patient return electrode. Unfortunately, due to the
inherent primary-secondary capacitance of isolation transformers, as well as the
construction properties of the generator, some stray leakage inevitably occurs. RF leakage
from an electrosurgical generator is of concern to users because an excessive amount can
place the patient, the surgeon, and other personnel coming in contact with the patient at
risk for an alternate site burn.
Almost all electrosurgical generators sold today have outputs isolated from ground to
prevent RF (radio frequency) leakage. By itself, an isolated output is not enough to
completely eliminate RF leakage. Other methods must also be used. However, some of these
methods can have a dramatic effect on the performance of the generator.
RF Leakage Current Measurement Setup |
 |
1. Supply mains
2. Table, made of insulating material
3. HF surgical equipment
4. Active electrode
5. Neutral electrode, metallic or in contact with metal foil of the same size
6. Measuring resistance, 200
7. HF current meter
8. Earthed conductive plane |
The potential hazard from RF leakage was initially recognized almost twenty years ago.
To minimize potential injuries, domestic and international safety standards were
developed. These standards require that the monopolar RF leakage be less than 150 mA
(0.150 Amps) as measured by a well-defined method. The value 150 mA was chosen to be less
than the minimum amount of RF current that could cause an alternate site burn given a
patient impedance of 200 ohms. The test uses the worst case scenario for RF leakage, which
is open circuit activation at the maximum power setting. There is a higher risk of RF
leakage with high voltage waveforms such as the fulgurate and spray coagulation modes.
In order to pass leakage standards, manufacturers use several methods of measurement.
Some electrosurgical generators are capable of detecting an open circuit condition and,
after a finite period of time, lower their peak output voltage. Since it is the peak
output voltage that initiates the sparking required for the fulgurate and spray
coagulation effect, using this leakage control mechanism could potentially degrade
generator performance. Due to the amount of time the generator needs to sense the open
circuit condition, momentary spikes of voltage can occur. As the surgeon moves
through gaps in the tissue or when the active electrode is withdrawn from the tissue, high
voltage spikes can momentarily cause the RF leakage current to exceed the
recommended specification.
To minimize RF leakage current, Valleylab generators modify the waveform duty cycle
that is, the amount of time the RF energy is on versus the time it is off. This
controls excessive RF leakage while retaining peak output voltage and the desired clinical
effect. By continuously monitoring the electrical system for open circuit or excessive
high voltage conditions using advanced electrical circuit techniques, Valleylab generators
achieve low RF leakage current while maintaining the desired surgical effect.
The monopolar cut and bipolar modes of the Valleylab Force FX with Instant Response
technology are handled differently from the monopolar coagulation modes. These Instant
Response controlled modes, by design, control RF leakage by the precise regulation of all
three output parameters: current, power, and voltage. Unlike generators from other
manufacturers, a separate RF leakage control system is not required to meet all of the
relevant standards requirements.
The technique used to minimize RF leakage should not compromise the ability of the
generator to meet the surgeons expectations. Due to its reduction in coagulating
affect, limiting the output voltage may not be the best method for RF leakage control. A
careful review of the methods used to reduce RF leakage is important for understanding
generator performance. The method of RF leakage control should emphasize both generator
performance and personnel safety.
Charles D. Allen has been involved in engineering research and development for over
20 years and currently coordinates electrosurgical product development for Valleylab.