Technical Aspects and Safety of Capacitive Coupling
Electrosurgery is the passage of radio frequency current through tissue to create a
desired surgical effect, such as coagulation or cutting. The use of radio frequency (RF)
energy (above 100 kHz) prevents neuromuscular stimulation while still providing excellent
tissue heating. Unfortunately, clinicians often do not fully understand the potentially
negative consequences of using radio frequency energy. These include RF leakage current
and high levels of capacitively coupled current that can cause alternate site burns. In
order to prevent capacitive coupling, it is important to understand how capacitors are
formed in the patient circuit, how capacitively coupled current occurs and what
preventative measures can be taken to reduce the likelihood of patient injury.
What is a capacitor?
A capacitor is an electric component that is capable of storing an electrical charge. It
is formed by two conductive plates that are separated by an insulating material called a
dielectric. Unlike a resistor that dissipates energy, capacitors store energy and return
it to the circuit in which they are connected. The amount of energy a capacitor can store
is measured in Farads. The formula for calculating capacitance is:

K = the dielectric constant of the insulator, Eo = a constant called the permittivity
of free space and is equal to 8.55 x 10-12 Farads/meter, A = the area of the plates in
square meters, and D = the distance between the plates in meters.
To purchase the full text of this article, click here...