Intra-Aortic Balloon Pump
In 1958 the idea of counterpulsation to treat left ventricular failure was introduced
by Harken. He theorized that the workload of the heart could be reduced and coronary
perfusion could be increased simultaneously.
By 1962 Moulopoulus, Topaz, and Kolff, working together at the Cleveland Clinic, had
developed the first intra-aortic balloon. Shortly thereafter Kantrowitz presented the
first intra-aortic balloon pump (IABP). In 1985 the development of the first prefolded
balloon occurred, greatly improving the process of placing the balloon.
The concept behind the IABP is counterpulsation, or, simply put, forcing the
oxygen-rich blood being pumped out of the heart to reverse course and flow into the
coronary arteries. This is achieved by inserting a balloon into the patients femoral
artery (in the groin area) and threading it into the aorta. Proper placement is to have
the tip of the balloon in the distal arch approximately 2 cm below the subclavian artery.
(If the balloon is too close to the heart, it can cause an obstruction in the flow of
blood to the brain.)
Once the balloon is in place, it can be inflated and deflated rapidly to achieve the
desired effect. With correct timing, the blood is forced into the coronary arteries,
supplying the heart with more oxygen. The deflation of the balloon prior to the next cycle
produces a lower pressure for the heart to pump against. Thus, the heart is receiving more
oxygen while performing less work. Balloon therapy is a short-term procedure designed to
allow the heart to gain strength and recover faster.
The balloon itself is made of polyethylene and is attached to a catheter. Several sizes
and lengths of balloons are available to accommodate various patients. Balloon designs
also vary. Some balloons have as many as three separate chambers. With this design, the
middle section is inflated first, and then the ends simultaneously inflate. Another design
has two sections, wherein the distal-end section inflates first to occlude the aorta, and
then the main body inflates, forcing a backward flow of blood. There are also
single-section balloons. Many balloons now have pressure transducers mounted at the tip to
give accurate arterial pressure readings to the operator.
The balloon is inflated with a carefully measured volume of gas, usually helium or
carbon dioxide. Helium, having the smallest molecules, gives the fastest
inflation/deflation times, while carbon dioxide is more readily absorbed into the body in
the event of a ruptured balloon, indicated by blood in the catheter.
The inflation and deflation of the balloon are timed to the cardiac cycle (generally
synchronized with the electrocardiogram) to give the greatest benefit and to prevent
backflow through the aortic valve. The ECG wave or the arterial waveform can be used to
determine the inflation and deflation point. The balloon should be inflated just after the
aortic valve closes. The dicrotic notch is the visual indication that the valve has closed
and can be used to set up the timing. The R-wave, indicating ventricular contraction, is
commonly used to trigger inflation. Deflation can be timed or triggered but must occur
before the next cardiac cycle.
Preventive maintenance includes verifying triggers and battery runtime, rebuilding
pumps, emptying water traps, and checking for pneumatics leaks. Inspection of the tank
yoke for damage , leaks and correct pin index should also be included. Remember your color
codes for medical gases and verify the correct tanks are being used. ( Helium tanks are
brown and carbon dioxide tanks are gray.) A knowledgeable technician using the
manufacturers specific procedures should perform preventive maintenance.
Common faults associated with IABPs include helium leaks, condensation problems within
the catheter, patient lead placement, and timing problems. Modern pumps have the ability
to trigger from internal ECG or arterial pressure waves, and many can also use external
sources of these waveforms. Occasionally the external signal will be delayed,"
causing incorrect inflation or deflation triggers. If the pump fails to trigger on the ECG
wave, the user should trigger the arterial waveform.
Occasionally a balloon will rupture, allowing the machine to draw blood into the
catheter, and if not detected, into the IABP itself. All parts of the pump that have
direct contact with the patient portion of the machine should be replaced if fluid
intrusion occurs.
| Questions |
1. What
color is a helium tank?
a. brown
b. gray
c. green
d. yellow2. Blood in the catheter
indicates_______________
a. normal operation.
b. rupture in the balloon.
c. condensation.
d. internal bleeding in the patient.
3. If the balloon is placed too high (too close to
the heart), what could happen?
a. The balloon would not inflate.
b. The blood flow to the patients brain could be obstructed.
c. The low helium alarm would sound.
d. The pump would have to run on batteries to avoid electrocution through the transducer.
4. If the pump will not trigger on the ECG wave, you
should______________________
a. manually set the cycle.
b. trigger from SPO2.
c. trigger from the arterial waveform.
d. trigger from the NIBP monitor.
5. IABPs are used to_____________________
a. increase cardiac perfusion.
b. lessen the workload of the heart.
c. pump blood throughout the body.
d. both a and b
Answers: 1a; 2b; 3b; 4c; 5d |
Glen L. Wolfe, CBET, CET, is lead technician at LaGrange (Ill) Memorial Hospital.
He has worked in the field for 12 years and holds an AA in biomedical technology. He is a
graduate of the US Army medical equipment repair school.