The radiation force balance principle is widely used for testing power output levels of
transducers. We present a step-by-step guide to the American Institute of Ultrasound in
Medicines recommended method for ensuring the reliability of your ultrasound probes.
A widely used method to test average power-output levels of therapeutic and diagnostic
transducers is the radiation force balance (RFB) principle. The American Institute of
Ultrasound in Medicine states that the radiation force balance was chosen for measurement
of ultrasound power because it is conceptually simple, relatively easy to build and use,
and provides a measurement which can be referred to NIST [National Institute of Standards
and Technology].
Several US companies manufacture commercial ultrasound power meters (UPMs) that utilize
the RFB principle. In this article the basic details are described. Other RFB devices may
vary in construction, usage, and calibration procedures.
This method is called RFB because the applied sonic force radiated from the
transducer-under-test (TUT) exerts a force proportional to the radiated sonic energy when
it is intercepted by the target. This force is measured on a mechanical or electronic
scale. The ultrasound power output in watts equals the radiation force in grams times the
velocity of ultrasound in water (for example, a power of 10 W corresponds to 0.68 g).
Figure 1.RFB Construction Details
1) test tank
2) sonic sound absorber
3) test media: degassed water
4) sonic target
5) coupling target to scale
6) transducer under test (TUT) positioning and support
7) scale to measure force exerted on target |
The Test Tank
A suitable test tank should have large enough water volume and solid mechanical
construction to allow accurate power measurement. A test-tank volume of 1 L was found to
be adequate. A 0.5 L capacity cannot support the sonic forces at high power levels; and
heat gradients, cavitation, or reverberation may cause unstable readings.
Sonic Sound Absorbers
The purpose of the liner in the test tank is to absorb sonic energy. The
effectiveness of the absorption is rated by the materials ability to dissipate
radiation and is measured in percent called reflection coefficient, which is the ratio of
the amplitude of a wave reflected from a surface to the amplitude of the incident wave.
Many different materials can be used to absorb sonic forces. Most of them are synthetic
rubber, such as neoprene. Some materials use glass air bubble beads with various levels of
different absorbing media to enhance the sonic absorption and diffusion.
Degassed Water
Most ultrasound standards specify water as the test media, since water approximates soft
body tissues and fluids in propagation velocity and can be readily distilled, filtered, or
degassed. Removal of some of the oxygenpreferably around 5 ppmis necessary for
accurate measurements. Small air bubbles cause secondary reflections and attenuation of
beam radiation force, which causes lower power readings. There are several ways to obtain
degassed water:
Boil tap or distilled water for 10 to 20 minutes in a mason jar without the lid
(suggested volume is a few liters). Cover the opening with plastic wrap and refrigerate.
Transfer (or siphon for minimum turbulence) to a 1-L CO2-type plastic bottle used for
carbonated beverages. The test tank for the UPM requires close to 1 L of water, so this
bottle size is preferred for one-time usage. To avoid oxygen buildup, refill the bottle
when storing.
In a hospital, obtain sterile distilled water from central supply and transfer
it to a 1-L CO2-type plastic bottle. Avoid using PVC or polyethylene containers, as over
time they will absorb oxygen.
Most drug stores carry 1 gal containers of distilled water. These containers
will also absorb oxygen over time, but distilled water has less oxygen than tap water.
Pull vacuum: Suction units provide adequate degassing. A rubber lid seal
connection is made between the vacuum source and the test tank or mason jar.
Sonic Target
All of the energy radiated from the TUT should intercept the 45° cone. A cone target,
similar to the one used by NIST, is milled from a 31/8-inch diameter solid aluminum rod to
an 80-mill wall thickness. The cone is sealed watertight, with the inside containing
either air or polystyrene foam. There are also absorbing-type targets covered with rubber,
but the most commonly used are the reflective cones.
Coupling Target to Scale
The sonic force exerted on the cone is transferred to a sensitive balance. Any
coupling method can be used as long as it does not degrade the accuracy of measurement.
TUT
The TUTs positioning and the radiated force path are shown in Figure 1. The
distance between TUT and the cone target should be kept to several wavelengths,
approximately 1/2 inch (at 1 MHz, wavelength is 1.54 mm).
Measuring Force Exerted on Target
To measure to the nearest 100 mW, scale sensitivity better than 0.01 g is required. The
gram readings are converted by the scale to read out directly in watts. A power of 1 mW
equals 68 µg.
Power Measurement Procedure for UPMs
It is important to select a test location free of vibrations and air currents
(manila folders can be used to reduce air turbulence). The UPM must be either placed on a
level surface, or leveled by using leveling knobs, if provided. Next, tilt the test tank
and pour the degassed water, with minimum turbulence, to approximately 1/2 inch from the
top of the tank. Then place the cone target into the tank. (Make sure that the cone is
uniformly wetted.) Position the TUT approximately 1/2 inch above the center of the cone
and tilt it at 45° into the water. (Make sure that the TUT surface is uniformly wetted.)
Allow a few minutes stabilization, then turn the power on, zero the balance, and verify
that the zero reading is stable. In this example, place the 1-g weight on the RFB
assembly; it should read 14.65 W. Turn on TUT power and take an average of three readings.
To reduce air absorption, cover the tank top between uses with cardboard or other
material.
Other Power Measurement Hints
Time duration of power measurements, particularly over 10 W, should be
kept short.
Extended test time increases water temperature and tends to start bubble formation,
causing attenuation and reflection.
Before new measurements, inspect for bubble formation on the target and on the
rubber absorber surface. Wipe off bubbles with a paper towel, tissue, or cotton swabs.
Keep TUT in the horizontal plane; tilting more than 30° will produce standing
waves as some of the energy bounces back from the water surface, increasing apparent power
output.
The electromagnetic restoring force of the scale may be affected by media, such
as heavy current power lines or instruments, generating strong magnetic fields.
The power output on therapeutic units tends to vary during warm up. Turn on the
unit with no power output for a few minutes to allow warm up before testing.
Figure 2.
 |
FDA Ultrasound Power Emission Accuracy
The US Food and Drug Administrations Radiation Control for Health and
Safety Act" specifies that the temporal average ultrasonic power be indicated on each
therapy unit within ±20% for all emissions greater than 10% of the maximum emission. A
commonly used calibration point is 10 W, with ±2 W maximum allowable deviation. The
measurement accuracy of the power meter is to be three to four times better, or ±5%. Most
power meters can meet these requirements.
Power Measurement Variations
Figure 2 shows changes in power output versus temperature from 40°F to 120°F test
conditions. Constant 10 W power is applied. Reference temperature is 77°F, frequency 1
MHz. Dissolved oxygen is a constant 6 ppm. Sonic properties of water are affected by
temperature as the density (mass of medium per unit volume) changes.
The higher the density, the more force is required to propagate. Compressibility
(fractional decrease in volume when sonic pressure is applied) also affects the speed of
ultrasound. The velocity changes with temperature: at 20°C, it is 1,480 m/sec in water,
but at 37°C, it increases to 1,570 m/sec. The highest density for water is at 3.98°C
(0.999 g/cm3) and the lowest is at boiling (approximately 100°C) where it is reduced to
(0.958 g/cm3). Therefore, it is best to use degassed water at ambient temperature rather
than refrigerated temperatures. The few tenths of a watt power output variations are
usually negligible and corrections can be made.
Denes Roveti is the technical director of OHMIC Instruments Co.