Troubleshooting
The American Radio Relay League (ARRL) defines electromagnetic interference (EMI) as
any unwanted signal or noise that distorts or disrupts a desired signal or an electrical
disturbance that interferes with a devices normal operation.
We have all talked about and seen much done about EMI that causes problems with medical
devices, usually as it relates to radio frequency (RF) transmissions. Most hospitals have
taken precautions when it comes to cell phones, but EMI can travel via other pathways.
According to ARRL, there is always a source for the interference, a device that is the
victim of the interference, and a transmission path that allows the interference to get to
the victim.
Stopping the interference at any of the three points prevents a device from becoming a
victim. If EMI is stopped at its source, it cannot spread. Removing the pathway prevents
EMI from getting to a device. Preventing EMI from entering a device stops that device from
becoming a victim.
EMI can travel through the air from a source to a victim by RF transmission. It can
also travel by conduction and induction. EMI travels by conduction when the source and
victim have a common conductive pathway. This pathway can be a wire or even conductive
tubing. Induction occurs when the source couples EMI to the victim through the pathway
provided by the proximity of their separate wires.
Since working on a special project to reduce EMI, I became more aware of EMI and its
prevention and include the possibility of problems due to EMI whenever I troubleshoot
equipment problems. I look more closely at how cables and power cords have been managed,
especially when I am unable to find a problem with a device after a user has reported one.
Following are some cable-management problems I found that allow EMI by either
conduction or induction to occur:
- Cords and cables mixed up and twisted with one another. The proximity of the cords and
cables to one another enabled EMI to occur.
- Cords plugged into outlets but not connected to any equipment. The unconnected cords,
wound up in a mass of cables, acted as a source of interference.
- AC outlet strips that caused interference on ECG monitors.
- Coiled cables and cords. The coiled wires set up fields that enabled EMI to occur.
I solved some device problems by correcting the above items. Therefore, I expanded my
EMI prevention program to include my areas of responsibility. Now, I look more closely at
cable management and take steps to prevent EMI problems.
I believe all technicians should pay much more attention to EMI prevention by following
simple cable-management procedures:
- Keep cords and cables as short as possible. Run them directly to each device.
- Route cords and cables away from one another. Make sure the cords and cables do not
become intertwined.
- Ensure cords are securely plugged into outlets and equipment and cable connectors are
securely connected at both ends.
- Disconnect and remove all unused power cords.
- Keep cords and cables away from other equipment and power sources.
- Replace suspected power strips.
- Do not coil excessive lengths of cords and cables. Instead, tightly bundle them and
secure them with tie-wraps, being careful not to create sharp bends that can crease the
cord and cable ends.
EMI problems can also be due to a buildings electrical system. Try using a
premium surge suppressor, not a cheap one. Periodically check the indicator lights and, if
the lights go out and remain out, replace the surge protector. Use an isolation
transformer to eliminate noise on an AC line. Get help from an electrician, who can check
electrical panels for loose or faulty connections and defective parts.
Remember to follow proper policies and procedures. Use medical-grade electrical
components when required.
Keep in mind that solving an EMI problem can be a recurring job. Equipment is moved or
replaced and renovations occurall meaning EMI can reappear.
EMI prevention is beneficial. If BMETs will take responsibly for proper cable
management, that effort will contribute to device uptime, patient well-being, and overall
customer satisfaction.
Ray Taft is senior BMET in the biomedical engineering department at Texas
Childrens Hospital in Houston.