Laboratory Equipment, Part 1
The clinical laboratory area of a hospital contains a wide variety of devicesfrom
the very simple heating block to automated chemistry systems.
In the simplest terms, laboratory equipment performs one or more of the following
functions on a sample:
Measures by using light, ions, conductivity,
weight, and color.
Separates by using electrical or rotational
force.
Counts cells, particles, and events.
Changes specimens by heating, cooling, drying
or reconstituting.
Sample preparation, delivery, and measurement are all functions that can have failure
points, but always consider the human factor before assuming that the device has failed.
Because of high capital costs and changing technology, many of the larger devices are
not owned by the hospitals but leased or obtained on rental basis, also called per-test
rental/lease. On non-hospital-owned equipment, service support is generally part of the
lease/rental agreement, so biomeds often only do incoming and visual inspections. In some
cases biomeds may provide a first look before calling the outside repair
service.
Clinical analyzersdiscrete, continuous flow, automatic, semiautomatic, or
manualperform six basic functions that cause problems.
The Basic Functions
Sample prep. The sample can be diluted, mixed with a reagent, or
filtered. This is often an automatic process where the samples are placed into special
trays or racks and loaded into the analyzer.
Sample introduction and transport. The prepared sample is aspirated or
injected into the analyzer, where it might be diluted further or have reagents added, and
moved to the detector chamber. In discrete systems this is done with an internal pipetter.
With a continuous-flow system, the samples are separated by air bubbles or water.
Detection. Detectors can be ion-specific electrodes, photometers, counters,
conductance measurements, or any combination thereof.
Processing. A microprocessor automatically compares and processes information
taken from the detectors to make calculations.
Flushing. Flush automatically cleans the transport system and detector so the
next sample is not contaminated by the sample just processed. On continuous-flow systems,
the flush is part of the sample-separation system. On some units there is an additional
flush that occurs after a set number of samples processed or time.
Display/printer. The results of each test on a sample are displayed on a CRT
or printed onto a paper format. Some devices use both methods. If the printer is external
to the device it may be part of the biomed service responsibilities.
Some devices also have the following parts:
System Connector. This communication port allows information to be sent directly
to medical records, billing, and other systems in the hospital. Information systems
personnel generally handle this interface.
Self-check/calibration. Automated systems have this function built in. It is
run one or more times per day/shift. When present, this is a valuable source of
information and needs to be checked before any troubleshoot is started.
External power conditioner or UPS. While major units are on dedicated power
circuits, some do require power conditioners to assure stability of the power. In some
hospitals a UPS may also be installed if the device cannot handle the power drop of 10
seconds or less between loss of power and the start-up of the emergency generators.
Blood Gas/pH Analyzer
With the widespread use of pulse oximeters, blood gas/pH analyzers in intensive
care units and operating rooms are not as common now as in the past. Even so, blood gases
are occasionally performed on patients. Blood gas samples are taken from arterial blood,
so the patient either has an arterial blood pressure line in or the authorized licensed
personnel has to perform an arterial stick. Because of the depth of an artery
in the body, this is a painful procedure. In addition, the stick may not clot quickly,
leaving the patient with a hematoma and associated bruising.
The arterial blood is drawn into a heparin-coated syringe that is placed in ice if it
is not to be analyzed within 2 to 3 minutes.
The sample is injected into the analyzer where it enters a constant-temperature chamber
(37°C) and is diluted. After temperature stabilization, it goes to the detectors, which
have been previously calibrated with buffer solutions, where it passes three electrodes.
These are ion-specific electrodes, one of which measures dissolved oxygen concentration in
the blood, one measures the carbon dioxide level, and one measures the blood pH. Once the
readings are obtained, the system flushes itself and injects a buffer to recalibrate the
pH electrode and prepare for the next sample. Calibration gases may also be injected at
this time to verify the accuracy of the other two electrodes.
The results of the tests are displayed, either on a CRT or as numbers on a digital
display, and in most instruments on a printed record. On some units, the results might be
electronically transferred to the patients medical record.
Common problems with blood gas/pH analyzers include out-of-date buffer solutions, no
cal gases, poor flushing, paper jams, not taking the time to properly warm up the
analyzer, and bad electrodes.
Biomeds should always do a self-test or auto calibration procedure before doing any
troubleshooting on the device. These steps will generally indicate where the first problem
is. After correcting the indicated problem, repeat the self-test and auto cal steps to
determine if the problem has been resolved.
Clinical Chemistry Analyzer
Clinical chemistry analyzers determine the concentrations of electrolytes,
proteins, metabolites, and drugs in samples of serum, plasma, urine, or other body fluids.
According to the ECTI comparison charts, depending upon the make and model of the device,
analyzers can perform between three and 255 preprogrammed tests.
In automated discrete systems, the fluid sample is placed into a rack or cassette where
it might be diluted or have reagents added and mixed with the sample. During this process
the samples could be warmed to match the temperature that is in the detection chamber, on
some devices. The samples are identified and the record started. Bar codes are generally
used to identify the sample and the patient from whom the sample came. The sample is moved
to the detection chamber. On some units only one sample at a time is tested, while on
others multiple samples are tested simultaneously. In cases in which the sample is carried
in a cassette, it is moved past the detectors in a sequence that allows for the treated
sample to be tested by the various detectors and with varying reagents. With this system,
scratching on the cassettes can cause problems for certain tests and the mechanical
positioning is critical.
More common is for the treated sample to be aspirated into the detection chamber, where
the ion-specific electrodes and spectrophotometry is performed. It may take several
samples to complete all tests, as different reagents could be used. After each sample is
tested the detection chamber is flushed before the next sample is aspirated. If the flush
is not total there can be carryover from the previous sample that could affect
the results. In some units the samples are divided and separated by water, so there is a
continuous flow into the detection chamber. The water used to separate the samples also
serves as a flush. As the tests are performed, the results are computed and compared with
normal limits before the results are printed/displayed. On some units any result outside
of a normal limit is printed/displayed in a different color. Units can be connected to the
hospital information system to send results back to the clinical floors/offices, to
medical records, and to billing.
If the report printer is not built into the device it may become a service issue in
some hospitals: Who takes care of the printer? Is it part of the contract on the device, a
separate biomed device, or part of information services?
On many of the clinical chemistry analyzers, power conditioners/UPS and water
treatment/filters may also be connected to the unit. Again there is a question of
ownership when it comes to servicing these external devices.
Common problems with clinical chemistry analyzers include mechanical positioning of
cassettes, loss of aspiration pump, inadequate flushing between samples, dirty detection
chambers, and paper jams.
The same principles and problems are present in all clinical analyzers
whether they do a few tests per sample or many. By following the same troubleshooting
procedures, biomeds can correct most problems without a making a service call to the
vendor. Also, be sure to use the self-test/calibration features found on most devices
before starting any additional troubleshooting. The problem is usually defined and the
remedy outlined in the service manual.
| Review Questions 1)
A blood gas analyzer generally has _____ electrodes.
a. 5
b. 3
c. 6
d. 2
2) Two common problems on automatic chemistry analyzers are
_________.
a. line power and water quality
b. poor flushing and dirty detectors
c. old samples and out-of-date reagents
d. user errors and power problems
3) Some clinical chemistry analyzers are capable of performing as
many as _______ preprogrammed tests.
a. 5
b. 25
c. 255
d. 500
Answers: 1-b; 2-b; 3-c |
David Harrington, PhD, is director of staff development and training at Technology
in Medicine in Holliston, Mass.
Contributing to this article is John Roy, a Technology in Medicine biomed assigned
to Womens and Infants Hospital in Providence, RI.