Laboratory Equipment, Part 2
In part 1 of this series, we looked at the basics of laboratory equipment, pH meters,
and automated analyzers. In this section we will cover more devices commonly found in
clinical lab settings.
Cell Washer
The cell washer/cell saver that is used in the operating room as part of an auto
transfusion system is different from the cell washer used in a laboratory. In the
laboratory the cell washer is sometimes confused with a centrifuge, [see ICC Prep, May]. A
cell washer looks and acts similarly to a centrifuge and has most of the same problems,
with several major differences:
There should be samples from only one patient at a time.
The whole blood has been previously spun down and the cells
decanted to a new tube.
The cells are mixed with 2%5% saline and spun again.
Three to four times additional normal saline (0.9%) is decanted
into the tube containing the cells.
The tubes overflow and the waste is collected in an external
chamber.
The saline rinses remove serum proteins from the surface of the cells so they do not
affect the tests that are performed when reagents are introduced.
Common Problems
Same as centrifuges
Problems with decanting pump for the saline
Overflowing waste container
Coagulation Analyzers
Coagulation analyzers range from the very simple clot timer, to semi- and fully
automatic devices that test all 12 coagulation factors. The automatic devices generally
cover both the hemostasisarresting of bleeding (clotting)and
homeostasismaintenance of blood fluidity. The machine may use whole blood, plasma,
or platelets for testing. Some units can test all three while others may be limited, so as
a biomed you have to determine what the medium is before doing any testing on the device.
Clot Timer
This small, often battery-operated device is commonly found in cardiac
catheterization labs and operating rooms. A tube with a whole blood sample is placed into
the well of the unit, which triggers both a small heater and a timer, (some older units
may have a start button that has to be pushed). On one side of the tube there is a light
and on the other a photo detector. As the blood clots, less light passes through the
sample, and, when fully formed, no light passes and the timer stops. This usually takes
less than 60 seconds.
Common Problems
Dead battery
Dirty light or photo detector
Defective switch
Automatic Coagulation Analyzers
These are generally large units located in a central laboratory area, sometimes
on reagent rental contracts. These units can use all three types of specimens and batch
testing is most common. The integrity of the sample must be assured to have accurate
results, if the samples are left unrefrigerated for any length of time they may start to
clot before any testing is done. Also, if the sample is drawn using too small a needle,
cells can be damaged and the sample compromised. (Blood should never be drawn for testing
with any needle less than 22 gauge in diameter the higher the number the smaller the
diameter.)
Samples are placed into a rack and introduced into the analyzer. This is a mechanical
process and alignment is critical for smooth movement. Once in place, various reagents are
added to the samples. The tests are done both by electrical conduction and photometric
detectors.
While sample collection is not part of what biomeds are responsible for, improperly
obtained samples will affect the results. If the majority of the samples are not getting
good results, look at the collection process. Samples to be tested in these machines must
be collected in plastic, siliconized, or nonwettable glass. Ordinary glass allows the
sample to coat the surface and will affect test results. Also investigate the size of the
needle used to obtain the samples.
Results are displayed in printed form and some units may be directly linked to a
computer that distributes result reports.
Common Problems
Mechanical alignment of the sample trays
Tray transport system in the machine
Reagent delivery problems
Wrong paper used in report generator
Waste product receptacle full.
Hematology Analyzer
Like coagulation analyzers, hematology analyzers range from very simple to fully
automated systems. Costs range from several hundred dollars to more than $100,000, with
wide variations in how many functions the units can perform.
Cell Counter
Very simply put, this is a hematology analyzer wherein whole blood sample is
injected into the unit and diluted and divided before being aspirated past optical sensors
that count cells. In the first sample division, the red cells are counted, the second
division counts white cells, and in the third division, platelets are counted. The results
are displayed on a screen or printed on paper. These units are used for quick counts for
patients in clinics and physician offices. The ability to accurately count platelets is
questionable in many of the simple units. Also these simple units do not adequately count
immature cells, particles or do any serum testing.
Though no longer in wide use, you might encounter a platelet counter in a clinic. This
works the same as a cell counter but was designed to count just platelets.
Common Problems
Dirty optics
Low suction levels so samples do not aspirate
Wrong paper in the printer
Tubing leaks, especially in the suction line
Incomplete flush between samples
Automated Systems
Most hospitals have automated hematology analyzers that combine the tests
performed by cell counters with some of the tests done by coagulation analyzers and may
include testing for certain proteins, serum, or particles.
Since these systems perform tests on samples from many patients, sample identification
is critical (this is also true of a hematology analyzer that has a loading carousel). The
samples are prepared and placed in racks that are transported automatically into the
system. The samples are diluted and aspirated into the system and separated from one
another by air, flush solution, or water. As samples pass from detector to detector,
reagents can be added automatically for specific tests. Test results are printed out with
patient identification. On some units, if a fault occurs on a testing step, all other
tests are stopped and cannot be restarted until the fault is corrected. On other systems,
the unit will keep running and just not perform the test where the fault has occurred.
Once the rack has been completed, the unit will not accept another test rack until the
fault has been corrected. You need to know how the system handles faults as it can affect
when service can be performed. On the automated systems, suction is generally obtained
from the hospitals central system. Central suction problems will affect a
units ability to aspirate the samples into the detectors. This is often the cause of
contaminated readings, as the flush solutions are incorrectly drawn through the system.
Water treatment systems should be considered part of the system when troubleshooting.
If water pressure drops due to clogged filters, the system may stop until the pressure is
returned to the correct levels. Mineral build up in the water system is becoming a problem
in many parts of the country. Flow rates and pressures will often change as the hospital
pipes get mineral buildup, requiring more frequent filter changes.
With all automated analyzers problems with the physical facility can and do cause the
devices to malfunction so you should be aware and check externally, for construction or
renovations, before getting into troubleshooting the units.
Common Problems
Bad or incomplete detector chamber flush
Loss of suction
Drop in water pressure
Voltage drops or spikes
Reagent or mixing pump failure
Misaligned racks on the transport track
| Review Questions |
1) How many coagulation factors are tested on an
automatic system?
a. 3
b. 5
c. 9
d. 122) How many times are blood
cells rinsed with normal saline in a cell washer?
a. 1
b. 710
c. 2
d. 34
3) Clot timers are most commonly found in
a. blood banks
b. hematology clinics
c. emergency rooms
d. operating rooms and cardiac cath labs
All answers are A. |
David Harrington, PhD, is director of staff development at Technology in Medicine
in Holliston, Mass.
Contributing to this article is Eric Massey, a TiM biomed at Women and Infants
Hospital in Providence, RI.