Tips on deciphering error codes, servicing, and performing preventive maintenance on
this challenging sterilizer system.
Like much of the equipment we biomeds service, the Steris System 1 sterilizer can
present quite a challenge. I must confess that I have never attended the System 1 service
school, and I am not an expert, but with the service manual in hand and an open phone line
to Sterisnot to mention an OR staff bearing down my backI have managed to
achieve some success.
My hope is that by sharing some of my experiences on the System 1, I will assist other
biomed techs in their approach to servicing the unit. I will also mention some valuable
service tips that I have picked up along the way, and I will conclude with my
recommendations for preventive maintenance of the unit. I recognize that experiences may
differ, and all service suggestions and recommendations reflect only my opinion.
Problems and Solutions
During the System 1s sterilize and diagnostic cycles, error codes are
generated that point the technician in the direction of a failure. However, the error
codes cannot be accepted at face value, and require interpretation. Here is an example
that occurred with one of our units. The error code water temp < 43°C was
generated (the machine specifications state that water inlet temperatures of
43°48°C are required). Seeing this error code, one might come to the conclusion
that the water temperature was actually out of spec. Not so fast. A look at the first few
steps of a properly functioning sterilize cycle should help us determine the problem:
- The chamber seal inflates by the activation of the air compressor. The chamber seal
reaches the desired pressure of 40 (+/- 2.5) psi, and the air compressor deactivates.
- Supply-inlet water runs through the unit and out to the drain to verify that its
temperature is within specification.
- When the supply-inlet water temperature is sensed to be within its acceptable limits,
the water begins to fill the units sterilize chamber.
In step 1, the air compressor can normally be heard turning on and then off as the
chamber seal reaches pressure. Regarding our error code of water temp
<43°C, however, the compressor never cut off. It could be heard continuing to
run. Nevertheless, the unit began to check the water temperature. There was one slight
problem: the supply-inlet water never flowed, because the chamber seal never achieved the
desired pressure. Consequently, the thermistor never measured the water temperature but,
in fact, measured the air temperature at a perfect 27°C. This is significantly less than
43°C, and hence the error code was generated. The solution? Replacing the leaking chamber
seal solved the problem.
Another scenario occurred with the same error code. In this failure, the air compressor
and chamber seal worked exactly as they should have. Even so, the water did not flow to
verify proper temperature.
Once again, the air temperature was measured, and the error code was generated. I am
aware of two possible causes for this lack of water flow. First, it is possible that the
prefilters A and/or B were dirty enough to totally block flow. Second, it is possible that
there was a problem in the float block, which could inhibit water flow.
The float block is mounted adjacent to the sterilize chamber and is used as an
indicator of the chamber water level. This is done via the float switch (LS3). Without
water in the float block, LS3 should read closed, indicating an unfilled chamber.
Conversely, when the float block is full, LS3 should read open, indicating that the
chamber level is full. Further, if the system controller thinks the sterilizer chamber is
full, the water inlet will not allow flow. Therefore, a problem with the float
blockor with the interpretation of its statuscan result in the water inlet
valve never opening. In our failure, the float block system was determined to be working
correctly. The solution? Replacing prefilters A and B solved the problem.
Occasionally during a cycle, water will pour out the front of the unit (definitely a
failure mode, and a good opportunity to refine ones housekeeping skills). There are
some common problems that can cause this to occur. Excessive buildup of water under the
equipment tray will push the tray up and off the seal, causing the water to spill out the
front. With a correctly functioning unit, after each cycle there is an accumulation of
50100 cc of water under the tray. Proper maintenance requires the operator to remove
this water before it becomes a problem. If there were to be more water than this after a
cycle, it would indicate excessive leakage. Its cause is often simply a problem with the
equipment tray (either a hairline crack, or broken/loose hose fittings). One should
inspect the tray carefully, especially at the rounded areas. A second common leakage point
is at the lid vent hole, which is glued to the bottom rear of the lid. It seems to be
somewhat fragile and will develop cracks. Unfortunately, if this is the case, the entire
lid requires replacement.
Another cause for a water spill is air trapped in the chamber. In a correctly
functioning unit, when water enters the chamber, the air in the chamber is displaced, and
passes to the float block, through the float block check valves (ck2 and ck3), and out to
the drain. If ck2 and/or ck3 fail, then the air is trapped in the chamber and will break
the lid seal, causing a spill. Problems with ck2 and ck3 are common, but fortunately, they
are not too difficult to access and replace.
Any article discussing common failures of the System 1 would be sorely lacking without
some mention of those infamous pinch valves (drain pinch valve [V5] and circulation pinch
valve [V6]). The pinch valves are soft rubber tubes, approximately the length and diameter
of an average forefinger. By applying air pressure to the outside of the rubber tube via
the air compressor, the pinch valve is squeezed closed, or pinched off. During pinch off,
water is blocked from flowing through the valve. By means of these valves, the fluid path
is directed as required. If not replaced periodically, these rubber valves will develop
small holes or tears and will not pinch properly. A variety of faults are possible with
the failure of one or both of these valves, and I will describe one such failure:
chamber level low.
During the circulation portion of the cycle, if the chamber loses enough fluid to cause
the float block switch to drop and become closed, a chamber level low error
can occur. In encountering a unit with this error code, I observed two symptoms. First,
during the circulation portion of the cycle, the air compressor never stopped running.
This indicated a leak somewhere in the pneumatic system. Simultaneously, water was exiting
the units drain port, until the chamber was empty. In a correctly functioning unit,
during circulation, V5 should be closed in order to block the water path to drain. Given
that this is accomplished by means of the air compressor, V5 was strongly suspect. In
fact, upon inspection, it was found to have a tear. This allowed air to be infused into
the circulation path to the chamber. Also, with the failure of V5 to become fully closed,
the path to the drain remained open. Solution? Replacing V5 solved the problem.
Service Tips and Suggestions
The following are tips and suggestions probably not found in the service manual.
Some of them are obvious.
- Do not be afraid of turning over the unit and removing the bottom cover. As a matter of
fact, get used to it. It is a good idea to get some assistance. When turning it over, lift
up the right side of the unit first, letting the excess water spill out to the left. This
will protect the electronics, which are on the right side.
- When servicing either the circulation pump, pinch valves V5 and V6, or the drain check
valve (CK8), gain access by removing the bottom cover.
- When servicing the high-pressure pump, gain access from the top (which is more
difficult).
- When replacing the high-pressure pump, also replace the high-pressure sense switch
(LS6).
- When removing the float block, it is extremely useful to drill a couple of small holes
in the side of the cover, to allow a screwdriver to gain access to the hose clamps.
Otherwise, the whole machine needs to be disconnected, turned over, and the bottom cover
removed. These access holes will not cause any problems, and will save much time and
energy. They can even be capped.
- When replacing the sterile water filter, soak it for 23 minutes in water prior to
installing. Otherwise, the unit may fail the diagnostic cycle two or three times before
passing (not helpful when you are attempting to repair a unit).
- Excessive water under the equipment tray (ie, >50100 cc) is most often caused
by a cracked or broken tray. The second most likely cause is a cracked lid vent hole.
- For a slow fill-time problem, try replacing prefilters A and B, as well as the sterile
water filter.
- If large bubbles are observed entering the chamber during fill, pinch valves V5 and/or
V6 are defective. The cycle may still pass, but have replacements available.
- When replacing one defective pinch valve, replace them both.
- Error code low heat rate will often be caused by a tripped heat sensor
switch, which is mounted to System 1s internal water heater. This can occur if the
incoming water temperature is excessively high. Remove the bottom cover, and check the
switch. Reset if necessary. If the incoming water temperature is too high, it will need to
be adjusted. At Union Memorial, Baltimore, we have 50-gal hot water heaters mounted in the
ceiling. This hot water is mixed with house cold water and fed into the System 1s.
- Make sure system 1s internal heater coil resistance is 10 ohms.
- Always take care to protect the printer from spilled water, or risk ruining the printer
board. (Believe me, I know.)
Recommendations for PM
My experience indicates that replacing the following parts on an annual basis
will keep real failure calls to a minimum (maybe even nonexistent).
- drain and circulation pinch valves (V5 and V6)
- drain check valve (ck8) rebuild kit
- float-block check valves (CK1, CK2, and CK3,) with O-rings
- chamber seal
- water filters A and B, sterile water filter, and the air filter
- sterilant aspirator
The total cost on these parts is approximately $1,000.
Additionally, while the bottom cover is removed, check for signs of water leakage.
Check the lid vent hole for cracks. Finally, check the functioning of the gas spring
cylinder, which holds the lid open, and replace it if necessary. Reassemble the unit,
reattach the plumbing, and test the unit by running both a diagnostic and regular
sterilize cycle. Then go tell the wonderfully patient nurse manager that all is well
because you have just completed a big-time PM on the System 1!
Summary
Learning how to service the Steris System 1 has been both challenging and
gratifying. Recently, my boss asked me whether or not I thought that the hospital needed
to purchase newer units. With much satisfaction, I was able to respond that as long as we
can continue to get parts, we can keep the System 1s up and running well.
Robert Resnicoff, CBET, studied biomedical engineering technology at Catonsville
Community College in Catonsville, Md, and has worked at Union Memorial Hospital, Baltimore
for the past 21 years.
Suggested Reading
Steris System 1 Processor Model 89A1/90B1 Service Manual (P/N 612001). Mentor, Ohio:
Steris Corp; 1994.
Steris System 1 Operator Manual (P/N 612025). Mentor, Ohio:
Steris Corp; 1994.