Infusion Pumps
Since the late 1960s, infusion pumps have been in widespread use in hospitals. All the
infusion pumps being presently marketed fall into one of the following three operating
systems:
- positive displacementincludes syringe and cassette pumps;
- linear peristaltica number of fingers milk fluid down a straight
channel; and
- roller or rotary peristalticwhere rollers push the fluid about 270° around a pump
head.
Some newer infusion pumps feature more than one channel, a drug
library, and/or a panel-lockout option to prevent unauthorized adjustments of pump
parameters.
At many hospitals, infusions pumps are the No 1 devices with which biomeds come into
contact. Many of the problems associated with infusion pumps are not technical in nature
but rather concern the users application of the technology.
Positive-Displacement Pumps
Positive-displacement pumps are probably the most common type of infusion pump in
use today. They employ a special cassette set with a built-in small chamber that is placed
on the pump. A small valve in the set, often called a duckbill, allows fluid
to fill the chamber during a pause in pumping. When pumping restarts, the valve closes to
force fluid toward the patient instead of back toward the bag.
The rate of flow is set by the speed of the pumping action and can range from less than
1 mL per hour to up to 1 L per hour. The flow rate contributes to the pressure used by a
pump to overcome resistance to flow. Today, most pumps will not generate pressure above
500 mmHg (millimeters of mercury), or 10 PSI (pounds per square inch). Some of the early
models could generate more than 50 PSI in pressure. This caused some major problems if the
intravenous (IV) needle being infiltrated went through the vein into soft tissue. Most
pumps now have built-in pressure sensors and allow for the setting of lower occlusion
pressures. The occlusion pressures should be checked during your normal preventive
maintenance (PM) cycle.
Other problems with cassette pumps, such as cost, slowed their acceptance for many
years. Once the cost problems were solved, their higher degree of accuracy made this type
of pump very popular.
The true syringe pump is the most accurate of all pumps, with some having accuracy on
delivered volumes to four decimal points. Cassette units are generally ±2% to ±5%
accurate, depending on the manufacturer.
Biomeds will encounter two types of syringe pumps in most hospitals. One type uses a
standard syringe, ranging from 5 cc to 60 cc in size. Not all pumps can be used with all
syringe sizes, which leads to the common problem of a clinical person trying to use a
syringe size or brand that is not compatible with the pump.
The travel of the plunger on the syringe is calibrated to the delivery rate in a
syringe pump (being a displacement pump). If a syringe has a larger diameter than what is
programmed for the machine, the same travel distance on the plunger will deliver more drug
to the patient. This problem is common when more than one brand of syringes is used in a
hospital.
The second type of syringe pump is the patient-controlled analgesia (PCA) pump, which
often uses a syringe specifically designed for that pump instead of a standard syringe.
Again, if the wrong size or brand of syringe is used, the delivered drugs could be either
overinfused or underinfused. There are generally two parameters that clinicians program
into the pump prior to starting the infusion. The first is the rate in mL/hr. The second
is volume to be infused (VTBI) in mL. Some hospitals may have a practice to set the VTBI
at approximately twice the rate. This means that the pump will sound an alarm in about 2
hours, indicating that the infusion is complete. The pump will then automatically switch
to the keep vein open (KVO) mode, which will infuse 1 mL/hr. This alarm alerts
the clinician to silence the pump, reset the pump, and check the patient to make sure the
infusion is OK.
Linear Peristaltic Pumps
Originally, linear peristaltic pumps used standard IV sets, had drop sensors, and
were not pressure limited. With some designs, it was possible to obtain more than 100 PSI
of delivered pressure. The majority of linear peristaltic pumps now in use are pressure
limited, use special sets, and have delivered accuracies close to that of the
positive-displacement pumps.
As the drive mechanism wears, accuracy can vary, especially at high occlusion
pressures. What happens is that the fingers do not completely close the
infusion set; and if the occlusion pressure of the patient is greater than the head
pressure from the container, some fluid is forced back to the container. One indication of
this problem is blood coming back into the IV line from the patient; another is
underinfusion of the solution. In either case, you have a problem.
One way to test for excess wear, from the bench, is to put 15 PSI of air pressure on
the patient end of the IV set and see if air bubbles work their way back to the IV
container. If there are air bubbles, you need to adjust the finger on the pump
to get total occlusion.
The delivered volumes for linear peristaltic pumps are the same as the cassette
positive-displacement pumps, and most have adjustments for the occlusion pressure.
Roller or Rotary Peristaltic Pumps
The roller or rotary peristaltic pump is not commonly used for IV drug
administrations but is more often used for pumping in feeding solutions for nutritional
support, such as total parenteral nutrition. The roller pump also is widely used for blood
in the operating room, during cardiac bypass, and in blood banks for processing or
separating cells from the plasma.
The roller peristaltic pump is probably the least accurate of all the pumps, and
calibrating the pump headsespecially for bypass pumpscan take time.
Some points to remember:
A controller basically counts drops; it does not pump fluid into a patient but
controls only the rate of flow. The maximum pressure generated is the height of the bag or
bottle above the injection site. Controllers cannot be used for arterial infusions.
The batteries on all battery-operated pumps must be replaced on a scheduled
basis. Generally, batteries should be replaced between 18 and 24 months. Allowing the
battery to fail before replacement may put the patient in danger.
Some of the newer pumps on the market feature more than one channel, allowing a
single pump to infuse up to four lines on the same patient. Certain pumps have a drug
library from which a clinician can tell the pump what drug it is infusing. Most pumps also
feature a panel-lockout option, which, when activated, disables the front panel. This
prevents the patient or anyone else from adjusting the pump parameters while it is
connected and infusing.
Pumps that do not have free-flow protection built in must be clearly marked, and
a training program must be in place for users. Most providers have addressed the free-flow
problem since the Joint Commission on Accreditation of Healthcare Organizations made it
one of the patient safety goals in 2002.
Pumps often suffer from sudden deceleration, or hitting the floor, and should be
mechanically inspected as part of the PM process.
A high percentage of the problems reported on IV pumps cannot be duplicated in
the shopoften indicating a problem with staff education. Work with staff-education
personnel to cut down on these calls. 24x7
Review Questions
1) Can a syringe pump use any syringe?
a. Yes
b. No
2) Most IV pumps, in general, have pressure limits of about
_____________________.
a. 50 mmHg
b. 50 PSI
c. 500 mmHg
d. 500 nanometers
Answers: 1-b, 2-c |
David Harrington, PhD, director of staff development and training at Technology in
Medicine (TiM), Holliston, Mass, is a member of 24x7s editorial advisory board.
Freeman Skip Sands is the TiM account manager at North Adams Regional
Medical Center, North Adams, Mass.