Most hospital staffers
understand that the M in MRI stands for
“magnetic,” and they know that magnets attract metal. They have
also heard about the disasters involving magnetic resonance imaging
equipment, such as patients being struck, pinned, or lacerated by a
ferromagnetic object pulled into the equipment’s powerful magnet.
Those objects have included oxygen bottles, tools, chairs, ladders, patient
lifts, traction weights, IV poles, laundry
carts, pens, hair barrettes, and scissors.But
that theoretical understanding can be as ineffective as a mother saying a
generic, “Be careful!” without real-world details to show
exactly how potent the forces at play really are. With careful training and
vigilance, you can have an accident-free MRI suite.
“I get a little upset when I find used needles
in the bore. That’s happened on quite a few occasions,” says
James Wilson, a Chatsworth, Calif-based senior MRI and computed tomography
(CT) specialist for the west zone imaging team at Master Plan Inc, a
national third-party independent service organization. The problem, he and
other experts explain, is not just simple human error; it’s
complacency. Even the best-intentioned technicians who use the suite day in
and day out will start to cut corners, save a few steps, and fall victim to
perhaps the most dangerous employee temptation of all: “Well, just
this once …”
The best way to overcome that is constant, relentless
employee education and as close to a
zero-tolerance policy as you can get. Actually, it is more like employee
“reminders” than employee “education,” Wilson
explains. The hospital staffers who use the equipment and service the suite
generally know that MRIs are complex, expensive, and potentially dangerous.
But that often remains a theoretical knowledge. Like high-rise construction
workers who know that a slip-and-fall could be deadly, those who work
around MRIs still use shortcuts when they know better—their version
of reaching just a little farther than they should for that connecting
beam, 35 stories in the air.
“I’ve been called in for four
accidents,” Wilson reports, “and all of them involved high-time
techs. The ones who know better are the ones I have to deal with.”
Indeed, he recalls, a staffer at one facility he manages even called to ask
about the safety of a device he wanted to take into the MRI suite and then
took it in even after he was cautioned against doing so. “That just
blew my mind,” Wilson says. “That’s why you have to keep
retraining. When I train with my team out in the field, they’re all
looking at me like, ‘Yeah, we know this.’ But I’m finding
that if I don’t keep harping on this, it does get dropped. We tend to
think we understand, but there are just so many issues beyond the magnetic
principles we deal with that it’s a constant education
issue.”
Magnetic Pull
The biggest and most important component in an MRI
system is the magnet. A horizontal tube, known as the bore of the magnet,
runs through the magnet from the front to the back. An MRI magnet is
rated using a unit of measure known as a Tesla (T). Another unit of measure
commonly used with magnets is the gauss (G) (1T = 10,000 G). The magnets in
use today in MRIs have been in the 1.5- to 2T range, but there is an
increasing availability of 3T MR systems for use in routine clinical
scanning.
When a patient is in the MRI scanner, the nuclei
(consisting of single protons) of his or her hydrogen atoms align with the
scanner’s magnetic field. Then, a second magnetic field, a radio
frequency (RF) signal, is applied, causing the aligned protons to move out
of their normal position. When the RF field is turned off, the hydrogen
nuclei realign back into their proper position. During the realignment
process, the nuclei send out signals. These signals are captured by the
computer system that analyzes and translates them into an image of the body
part being scanned.
There are, explains Glen Wolfe, CBET, CET, manager,
biomedical services, at Trace Services Inc, Elk Grove Village, Ill, three
basic types of MRI units: superconducting, permanent, and resistive
magnets. In the first two, a magnetic field is present at all times once
the machine is installed. The resistive magnet is turned on and off as
needed.
Respecting the Power of MRIs An MRI suite can turn into very dangerous place if strict precautions are not observed. Metal objects can become projectiles, small metal objects can be pulled out of pockets and off the body without warning, and they can fly toward the magnet’s opening at high speeds, posing a threat to everyone in the room. Most MRI systems also erase credit cards as well as anything else with magnetic encoding. The magnetic force exerted on an object rapidly increases as it nears the magnet. For personal safety and for the safety of everyone else in the room, exercise caution and remember:
•Radio-frequency signals can interfere with many electronic devices outside the MRI machine, Glen Wolfe, CBET, CET, manager, biomedical services, at Trace Services Inc, Elk Grove Village, Ill, points out, which can cause them to work erratically or stop working altogether. Insulin pumps have been known to stop working in the presence of an MRI.
•John Engel, CRES, a retired senior systems training specialist at DITEC, Solon, Ohio, also notes that pacemakers can be reprogrammed. “That’s not,” he says, “something the average heart patient needs.” He adds that he has a piece of metal in one of his fingers that he can feel moving in the presence of an MRI.
•Even a large dumpster placed outside one scan room caused image distortion.
•Superconducting coils in the device are cooled with liquid helium and/or nitrogen, Wolfe adds. When a service technician refills the liquid, the exposure could cause a finger or hand to freeze instantly.
•If the magnet’s coil gets too hot, he continues, the liquid turns to a gas, which builds up pressure. Venting should be directed outside the building, but if a rapid gas leak vents into the scan room, oxygen is displaced with helium—and asphyxiation can occur.
•ECRI notes that a sandbag placed in the vicinity of one patient’s groin for site compression and control of bleeding was violently pulled from its position on the patient’s body and became pinned to the wall of the MRI tube as the patient was brought toward the machine. The patient received bruises to her chest and head and, although the MRI unit sustained no damage, two men were required to remove the sandbag from the bore.
•Nonmedical metal objects are subject to the magnetic forces as well, Engel notes. “Once you’ve lost a couple sets of credit cards,” he says, “you get the idea.”
•James Wilson, a Chatsworth, Calif-based senior MRI and computed tomography specialist for the west zone imaging team at Master Plan Inc, also reports that welders may have tiny shards of metal in their eyes, and if they enter the MRI suite, blindness can result.
•Engel points out that tools can become deadly, too. “I very nearly got stabbed in the throat with a screwdriver because somebody had mixed it in with the nonmagnetic tools,” he says.
• To make his point about the power of MRIs, Wilson uses two common items in his staff-training programs. “I show them an aluminum ladder being affected by the magnet,” he says, “and I show them a ball of steel wool encapsulated in plastic, too.” —RJ
Minimize the Risks
To keep up with the latest technology in the MR field
and to increase MRI safety, ongoing employee training is essential, making
sure underlying policies and procedures are in place, too.
“Most accidents aren’t caused by
technicians or biomedical engineers, but by outside vendors,” Wilson
explains. “They should have to check in so we can do the screening. Some hospitals are good at saying, ‘I don’t care
if you’re the MR service engineer; I’m going to screen you myself.’ Now, the rest have to step up and
ask, ‘How do I minimize the risk?’ ”
Eight practices a facility can implement right now to
help ensure a safer MRI suite are:
• Restrict access with a screening sheet that
requires people going into the suite to sign it, saying they have no metal
on their persons. Also, many institutions require someone inside the suite
to unlock the door, while others restrict access altogether unless a
patient is being prepped for a procedure. “Whoever is in
charge,” Wilson recommends, “maintains the key and says,
‘You’re not going in there unless I screen you.’ ”
• Be sure an oxygen monitor is installed in the
room, and convert all O2tanks in the hospital to an MRI-safe type.
• Store a nonferrous tool kit—for use only
by biomedical engineers and only in the scan room—in the MRI computer
room.
• Allow only a verifiably trained engineer to
work on MRI equipment.
• Adopt a policy requiring all equipment
purchases for the MRI suite to be reviewed and approved by the biomedical
engineering department.
• Clearly mark MRI-safe equipment. One hospital
paints oxygen tanks, gurneys, and wheelchairs bright yellow to identify
them as MRI safe.
• Invest in the type of walk-through metal
detector found in airports.
• Do not rely entirely on safety nomenclature,
such as “MR Safe,” “MR Conditional,” and “MR
Unsafe,” from the various national organizations. “There should
only be one marking: MR Safe,” Wolfe says. “All the rest are
confusing. Either it is safe or it is not.”
Costly Equipment Damage
MRI risks extend beyond human injuries. While patient
safety is, of course, paramount, equipment damage can also occur in an MRI
accident. A ferromagnetic object lodged deep inside a tight MRI bore may
not only be hard to reach, but generating enough force for its removal may
prove difficult, resulting in extensive downtime. In an emergency, the
magnet may need to be turned off quickly, putting the system out of service
for days, entailing substantial financial losses, which can exceed $500,000
if the magnet has to be replaced. MRI accident reports have also detailed
dislocation of and other damages to the magnet, as well as damage to the
gantry, fan blower, and body coil antenna.
Indeed, Wilson stresses, he knows of one site that had
not even opened its brand-new MRI to patient use and was down $140,000
because a janitor was buffing the floor around the magnet and the magnet
sucked the buffer inside. In addition, technicians who think they know what
they are doing can make matters worse when they try to remedy such
situations. “Don’t try to hide a mishap from us,” he
emphasizes. “Part of the damage is obvious. But part is what you
can’t see.”
And the risk is just going to get worse. In fact,
Wilson points out, “With 3T magnets, the advanced type that generate
3 Teslas of force, it becomes a physiological issue. I know of an engineer
who was hospitalized just because he was in the room too long working near
the magnet. We didn’t have to deal with that even 2 or 3 years ago.
And we haven’t hit a saturation point in the field, either. There are
more and more MRI machines in departments that haven’t had them and
where the facilities’ staffers aren’t adequately
trained.”
John Engel, CRES, a retired senior systems training
specialist at DITEC, Solon, Ohio, notes that biomedical engineers
themselves also need to step outside of their day-to-day mind-sets and
focus anew on MRI safety. For one thing, he stresses, “Whoever
manages the facility needs to invite the local fire department in and show
the firefighters how to quench the magnet.” That, of course, should
happen only when a life or limb is threatened, Wilson adds. “The
signs on some MRIs’ quench buttons read, ‘Quenching the magnet
is quick and easy. Ramping the magnet back up is long and costly. Are you
sure you want to do this?’ ”
And, Engel adds, staffers need to battle the
complacency that leads to most accidents. “I use the example of a
group of experienced friends who planned to videotape one of them
making a parachute jump. But they forgot the parachute. That means one of
two things. Either nobody liked the guy, or they all had the,
‘He’s an expert; he knows,’ syndrome.”
Russell A. Jackson is a contributing writer for 24x7.