Mammography is the
examination of breasts with x-rays.
Surprisingly, mammography has a history that dates back to the early 1900s.
In 1913, Albert Salomon, MD, from the Surgical University of Berlin was the
first to document the radiographic evidence of a tumored breast.
Salomon used glass plates and early x-ray technology to create his
images.
Due to poor image quality, advances in mammography
were halted until the early 1950s. Using microcalifications, Raoul
Leborgne, MD, realized the need for perfect technique when studying
mammography. He noted four major points: the need for a low kilo voltage
peak (kVp) technique, the need for collimation of the x-ray beam, the need
for high-contrast images, and the need for tissue compression. In 1956,
Robert Egan, MD, noted ways to achieve
Leborgne’s equipment needs by optimizing
the x-ray equipment, using specialized film processing, and implementing specialized training for mammography
radiologists.
In the 1960s, Professor C.M. Gros made advances in the
technology and techniques used in mammography. First, he replaced the
tungsten target tube with molybdenum, which increased the breast’s
image contrast. Also, he noted that vigorous compression is needed for
quality images, more so than Leborgne thought. Gros, along with the
Compagnie Generale de Radiologie of France, developed the first dedicated
mammography x-ray machine: the Senographe.
The Senographe had significant changes from a general
x-ray machine. These changes included: a rotating tube stand—built
for positioning; the x-ray tube was cooled with molybdenum and water;
the x-ray tube exit window material was changed to beryllium and replaced
the glass previously installed, allowing the low-energy photons to
exit, which is crucial for mammography; the
focal spot was reduced from 1.5 to 2.0 mm down
to 0.7 mm; a dedicated control panel was created to allow low-dose
selection by the operator; a variety of interchangeable cones were created
for greater collimation; and a compression paddle was added to the tube
stand.
Compression force has to fall within the range of 25
to 40 pounds. Keep in mind that the most common problems with compression
is the cracking of the paddles and the force applied. When cracks start to
develop on the paddles themselves, these cracks can show up on the
processed x-ray film. If compression force starts to decrease, image
quality can be affected. This can usually be adjusted by a potentiometer on
the tube stand itself or on a PCB inside the machine’s generator.
Imaging Facts
kVp measures the force of electrons that hit the
target and determines the x-ray’s penetration force. A low kVp
generates longer wavelengths, which are best for imaging soft breast
tissue. About 25 kVp is the norm for mammographic imaging if a grid is not
being used. If a grid is in place, you can typically add 2 kVp for the
technique. Ranges below 25 kVp will increase radiation dose, and ranges
above 30 kVp will compromise image contrast.
Milli amperage seconds (mAs) is the current driven to
the filament of the x-ray tube cathode. The mAs regulates the amount of
electrons available to penetrate the target. Also, the mAs is variable,
allowing the technologist to change settings during an exam to get the best
possible image for that patient. Generally, the kVp is a set number and is
not changed.
Automatic exposure control (AEC) is a circuit that
monitors the x-rays that hit the image receptor under the film cassette and
automatically stops the exposure at a preset value—known as
photo-timing. When AEC is used, only the mAs value is changed to create a
consistent film density throughout the image.
With the introduction of digital systems for
mammography, we can expect some radical changes in equipment, techniques,
and storage of the images. Digital systems have been shown to be better on
younger patients and those with very dense breasts. There is some
resistance to digital systems—mostly due to their price—but
digital systems are coming and will replace most of the units presently in
use within 5 years.
Xerography is a dry photographic or photocopying
process in which a negative image (formed by a resinous powder on an
electrically charged plate) is transferred to and thermally fixed as
positive on paper or other copying surfaces. It was first introduced to the
medical field in the 1960s with mammography, and it used two units: a
conditioner and a processor. First, a photo-conducting,
selenium-coated aluminum plate is electrically charged and placed in a cassette, which completes the
conditioning process. After the cassette is
exposed to an x-ray beam, a dormant transparent image is existent on the
plate. The plate is then removed from the cassette and placed in the
processor. The dormant image is now exposed
to a blue powder of charged particles. The powdered image is then
transferred onto a plastic sheet. When the plastic sheet is heated, the
powder image becomes part of the plastic and is then visible.
Mammographic view boxes have specific light-output
regulations for viewing the processed film. View box light intensity is
measured in nit—this is a unit of visible-light intensity. One nit is
equivalent to one candela per square meter, and a candela is equivalent to
1 candle power. A mammographic view box must obtain a minimum output
strength of 3,000 nit. This value can be tested with a calibrated natural
daylight meter.
Assuring Quality
Daily tests must be performed to assure the quality of
the x-ray image generated on the film and of the processing unit of the
film. Temperature readings of the developer and fixer must be checked daily
and must fall within acceptable ranges, per the processor’s
manufacturer. These values depend on the brand of developer and fixer being
used, but they are generally between 33° and 35° centigrade. A
phantom is an acrylic block used as a control and simulates a patient. This
image is then processed, and the exported film is then tested with a
densitometer. A densitometer is an instrument for determining optical or
photographic density of film. The optical density of the center of the
phantom image must be between 1.10 and 1.50 for films exposed at the kVp
used by the facility. A sensitometer—an instrument for measuring the
sensitivity of photographic material—is then
used. A sensitometer will expose the film to a controlled light to be
processed for later review. When the film is processed, it is then tested
on the densitometer to determine the range of optical densities that
provide optimal film contrast. When checked, these values must not exceed +
or – 0.15 of the previously checked optical density with the phantom.
These tests should be performed by the technologists, not the BMET.
David Harrington, PhD, is director of staff
development and training at Technology in Medicine (TiM), Holliston, Mass,
and is a member of 24x7’s editorial advisory board.
Carl Genereux is a TiM account manager at Marlborough
Hospital, Marlborough, Mass.