I have a Voluson 730 pro that reboots in mid scan. Any ideas on what it could be?

Check the error logs and the status of the power supply LEDs on the back of the GEF box when the problem occurs. It can be a software corruption. Make sure all patient data has been transferred to the hospital DICOM server, then create a recent system backup including system options. Reload system software according to the procedure in the service manual.